Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 105
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.
To edit the name of a data channel, follow the above key sequence and then press:
From the end of the previous key sequence …
SETUP X.X
NAME:CONC
C O N C - - ENTR EXIT
ENTR
accepts the new string
and returns to the previous
menu.
EXIT
ignores the new string
and returns to the previous
menu.
Press each key
repeatedly to cycle through the
available character set:
0-9, A-Z, space ’ ~ !
©
# $ % ^ & * ( ) - _ = +[ ]
{ } < >\ | ; : , . / ?
SETUP X.X
NAME:CONC
<SET SET>
EDIT
PRINT EXIT
6.12.2.3. Trigger Events
To edit the list of data parameters associated with a specific data channel, press:
SETUP X.X EVENT:ATIMER
<PREV NEXT> ENTR EXIT
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
From the
DATA ACQUISITION
menu
(see Section 6.12.2.2)
Press each
k
ey
repeatedly to cycle through the
list of available trigger events.
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X EVENT:ATIMER
<SET SET> EDIT PRINT EXIT
ENTR
accepts the new string
and returns to the previous
menu.
EXIT
ignores the new string
and returns to the previous
menu.
Exits to the Main