INSTRUCTION MANUAL
MODEL 401
O
3
CALIBRATOR
© TELEDYNE INSTRUMENTS
ADVANCED POLLUTION INSTRUMENTATION DIVISION
(T-API)
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
TOLL-FREE: 800-324-5190
FAX: 858-657-9816
TEL: 858-657-9800
WEB SITE: www.teledyne-api.com
01124
REV. J2
Copyright 2005 Teledyne Advanced Pollution
Instrumentation
11 July, 2005
Vedere la pagina 0
1 2 3 4 5 6 ... 113 114

Sommario

Pagina 1 - CALIBRATOR

INSTRUCTION MANUAL MODEL 401 O3 CALIBRATOR © TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) 9480 CARROLL PARK DRI

Pagina 2 - SAFETY MESSAGES

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 The warranties set forth in this section shall be of no force or e

Pagina 3 - TABLE OF CONTENTS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 10-4

Pagina 4

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 11 MODEL 401 SPARE PARTS AND EXPENDABLES KITS 11.1 Spare Parts Fo

Pagina 5

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 11-1: Spare Parts For CE Mark Units (Continued) Part No.

Pagina 6

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 11.2 Spare Parts For Non-CE Mark Units Table 11-2: Spare Parts F

Pagina 7 - LIST OF FIGURES

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 11-2: Spare Parts For Non-CE Mark Units (Continued) Part N

Pagina 8 - LIST OF TABLES

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix A LIST OF AVAILABLE MODEL 401 OPTIONS Table A-1: Availa

Pagina 9 - 1 INTRODUCTION

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK A-2

Pagina 10 - 1.3 Principle Of Operation

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix B RECOMMENDED ZERO AIR SYSTEM The following is from the

Pagina 11

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK B-2

Pagina 12 - 1.3.1 Interferent Rejection

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix C REFERENCES 1. Calibration of Ozone Reference Methods,

Pagina 13 - 1.4 Specifications

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE Model 401 Photometric Calibrator is designed to be a dedicat

Pagina 14

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK C-2

Pagina 15 - 1.4.12 Weight

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 D-1 Appendix D SOFTWARE MENU TREES Figure A-1: Model 401 Sampl

Pagina 16

ne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 D-2 Figure A-2: Model 401 Setup Software Menu Tree Teledy

Pagina 17

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Appendix E ELECTRICAL SCHEMATIC INDEX Table E-1: Electrical Sche

Pagina 18

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK E-2

Pagina 19

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 The intensity of light is converted into a voltage by the detector

Pagina 20

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.4 Specifications Photometer Ranges User selectable to any full

Pagina 21

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.4.1 Noise (PPB RMS) The standard deviation of reported concentra

Pagina 22 - 1.6.2 RS-232

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.4.10 Flow Rates 800 scc/min ± 10% (standard cubic centimeters/mi

Pagina 23

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE See Figure 1-4 for rear panel pneumatic connections. Span gas

Pagina 24 - 1.6.5 Output Flow Adjustment

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10. Turn on the M401 by switching the switch on the lower right co

Pagina 25

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 14. Generate the Zero Air. a. Press ZERO from the main menu and e

Pagina 26

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-1: Model 401 Ozone Calibrator Block Diagram 1-11

Pagina 27

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 SAFETY MESSAGES Your safety and the safety of others is very impor

Pagina 28 - 1.7 Operation Verification

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.7 Figure 1-2: M401 Rear Panel Electrical Connections 1-12

Pagina 29 - Table 1-1: Test Values

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-3: Rear Panel Electrical and Pneumatic Connections (CE

Pagina 30 - 1.8 Options

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.6 Electrical And Pneumatic Connections 1.6.1 Electrical Connecti

Pagina 31 - 2 OPERATION

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-4: RS-232 Pin Assignments 1-15

Pagina 32 - 2.1.7 Password Protection

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.6.3 Pneumatic System The Model 401 Photometric Calibrator is equ

Pagina 33

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-5: M401 Calibrator Pneumatic Connections 1-17

Pagina 34 - 2.2 Front Panel

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-6: M401 Rear Panel (External Dry Zero Air) 1-18

Pagina 35

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 1-7: M401 Calibrator Chassis Layout 1-19

Pagina 36

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.7 Operation Verification The Model 401 Calibrator is now ready f

Pagina 37

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.7.1 Final Test And Calibration Values Table 1-1: Test Values Tes

Pagina 38

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 TABLE OF CONTENTS SAFETY MESSAGES ...

Pagina 39 - 2.2.3 Status LEDs

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1.8 Options 1.8.1 Rack Mount With Slides This option, including sl

Pagina 40

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2 OPERATION 2.1 Key Features The important features of the Teledyn

Pagina 41 - 3 CALIBRATION

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.1.5 Data Acquisition The Teledyne API Photometric Calibrator con

Pagina 42

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-1: Password Levels Password Level Functions Allowed No

Pagina 43 - 3.6 Photometer Calibration

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.2 Front Panel This section describes the operator interface from

Pagina 44

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-2: System Modes Mode Meaning STBY xxxx (1) Standby mode

Pagina 45 - 3.9 Hold Off

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-3: Test Measurements Test Message Meaning TIME=xx:xx:xx

Pagina 46

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 2-4: Warning Messages Warning Message Meaning SYSTEM RESET

Pagina 47 - 4 SETUP MODE

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 2-2: Illustration Of Normal Display If TEST is pushed, th

Pagina 48

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.2.3 Status LEDs The three status LEDs to the right of the displa

Pagina 49 - 4.7 Setting The Date

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2.1.2 O3 Analog Output...

Pagina 50 - Concentration Range

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 2-10

Pagina 51

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3 CALIBRATION The M401 is designed to be used as a dedicated ozone

Pagina 52 - 4.15 Software Configuration

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3.3 Automatic Zero/Span (or Multipoint) Check Automatic zero/span

Pagina 53 - 0 mV 1000 mV

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3.5 Summary Of Front Panel Calibration Controls The calibration co

Pagina 54

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 3. Allow to sample zero gas for approximately 10 minutes. 4. Fol

Pagina 55 - 5 DIAGNOSTICS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE Teledyne API recommends that contact closures remain closed a

Pagina 56 - 5.2.1 D/A Output

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 3-6

Pagina 57 - 6 HANDLING WARNINGS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4 SETUP MODE This section describes the setup variables which are

Pagina 58

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.4 Examining The Ozone Formula Slope And Offset The slope and off

Pagina 59 - 7 RS-232 COMMUNICATIONS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.5 Setting the O3 Generation Mode There are three modes of feedba

Pagina 60 - 7.1 DAS Reporting

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 5.2.1 D/A Output...

Pagina 61 - 7.2 Warnings

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.8 Adjusting The Clock Speed In order to compensate for clocks wh

Pagina 62 - 7.3 Status/Control

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.12 Setting The Analyzer I.D. Each Analyzer may be programmed wit

Pagina 63

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 4.15 Software Configuration The software configuration can be disp

Pagina 64 - 7.4 Diagnostics

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 4-1: Setup Variables Button Sequence Function Default Li

Pagina 65 - 7.5 Test Measurements

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 4-8

Pagina 66

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 5 DIAGNOSTICS The Teledyne API ozone analyzer contains two levels

Pagina 67

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 5-1: Test Channel Output Test # Name Purpose 1 NONE No ou

Pagina 68

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 6 HANDLING WARNINGS When a system warning occurs, a warning messag

Pagina 69 - 8 ADJUSTMENTS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 INTENTIONALLY BLANK 6-2

Pagina 70

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 7 RS-232 COMMUNICATIONS The Model 401 Photometric Calibrator featu

Pagina 71

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6.4 Checking The DC Power Supply Board...

Pagina 72 - 8.4 Flow Readout Adjustment

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 The "X" indicates the message type as shown above in Tab

Pagina 73

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 7.2 Warnings Whenever a warning message is displayed on the displa

Pagina 74 - 8.6 CPU

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-2: Warning Message Clear Commands Command Warning Messag

Pagina 75 - 8.7 Status Lines

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 To do a remote adjustment via the RS-232 interface, the host compu

Pagina 76 - ± 500 mV

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 7.4 Diagnostics The diagnostics mode can be entered from the RS-23

Pagina 77

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-6: Special Diagnostic Commands Command Function "D

Pagina 78 - 8.12 Bench Feedback Option

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-8: Test Measurement Request Commands Command Test Measur

Pagina 79 - 9 TROUBLESHOOTING

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 For example, to see the analyzer lamp temperature set point, the h

Pagina 80

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 7-9: RS-232 Variable Names Var. Name Button Sequence Lega

Pagina 81 - 9.2.3 Checking The Keyboard

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8 ADJUSTMENTS NOTE The operations outlined in this chapter are to

Pagina 82 - Table 9-1: Warning Messages

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 LIST OF FIGURES FIGURE 1-1: MODEL 401 OZONE CALIBRATOR BLOCK DIAG

Pagina 83

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 8-1: Electrical Block Diagram 8-2

Pagina 84

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.2 A/D - D/A Calibration Procedure Due to the stability of modern

Pagina 85

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Next, to automatically calibrate all the DACs, press SETUP-MISC-D/

Pagina 86 - 9.5.6 Cannot Span

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Figure 8-2: Flow And Pressure Readout Adjustment 8-5

Pagina 87 - Components

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.5 DC Power Supply Overall performance of the DC power supply may

Pagina 88

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.7 Status Lines Table 8-1: The Calibrator Has The Following Stat

Pagina 89 - 9.6.3 Checking The V/F Card

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 CAUTION UV light present. Do not pull the lamp from the optical b

Pagina 90

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 8.10 Ozone Generator Calibration The ozone generator can be calibr

Pagina 91

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 To view the current dark offset, press SETUP-MISC-DARK-VAL. Press

Pagina 92 - ° C=2.5V, ± 0.125V/°C

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9 TROUBLESHOOTING NOTE The operations outlined in this chapter ar

Pagina 93

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 LIST OF TABLES TABLE 1-1: TEST VALUES...

Pagina 94

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 If these actions all occur, it is probable that the Calibrator&ap

Pagina 95

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.2.2 Checking The CPU And Display When the analyzer is turned on

Pagina 96 - TELEDYNE

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 9-1: Warning Messages Warning Message Meaning Corrective

Pagina 97 - 10 ROUTINE MAINTENANCE

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.4 Troubleshooting Using Test Function Values The Model 401 Phot

Pagina 98 - 10.3 Cleaning Of Sample Cell

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 9-2: Test Function Values (Continued) Test Function Meanin

Pagina 99 - 10.4 Changing The Prom

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.5.1 Noisy Or Unstable Readings At Zero 1. Check for leaks in t

Pagina 100 - INTENTIONALLY BLANK

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.5.4 Analog Outputs Do Not Agree With Front Panel Readings 1. C

Pagina 101 - EXPENDABLES KITS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6 Troubleshooting Individual Sub-Assemblies And Components The

Pagina 102 - Teledyne API Model 401 O

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6.2 Troubleshooting Temperature Problems The Model 401 has been

Pagina 103 - (table continued)

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Chassis Temperature Turn the analyzer off and remove the DC Power

Pagina 104

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 1 INTRODUCTION 1.1 Preface Teledyne API is pleased that you have p

Pagina 105 - OPTIONS

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 2. Confirm that all jumpers on the V/F board are set properly, a

Pagina 106

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9.6.4 Checking The DC Power Supply Board A schematic and physical

Pagina 107

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Four Temperature linearization circuits are contained on the DC P

Pagina 108

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Table 9-5: UV Source Lamp And Detector Diagnostics UV Source Lam

Pagina 109 - Appendix C REFERENCES

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 NOTE In cold, ambient conditions, it may require 5 to 15 minutes

Pagina 110

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 9-17 WARNING Hazardous voltage present - use caution. 1. Unplug

Pagina 111

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 Warranty/Repair Questionnaire TELEDYNE INSTRUMENTS Advanced Po

Pagina 112

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10 ROUTINE MAINTENANCE NOTE The operations outlined in this chapt

Pagina 113

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10.2 Leak Checking There are two methods of leak checking: 1. By

Pagina 114

Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual, 01124, Rev. J2 10.4 Changing The Prom 1. Locate the CPU card by referring to Fi