Teledyne 200A Manuale Utente Pagina 142

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 225
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 141
Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
8.8 Pneumatic Line Inspection
Particulate matter in the pneumatic lines will affect both flow rate and response time. It is
important that the pneumatic system be periodically inspected and thoroughly cleaned if
necessary. Clean by disassembling and passing methanol through three times. Dry with nitrogen
or clean zero air.
Also inspect all pneumatic lines for cracks and abrasion on a regular basis. Replace as necessary.
Refer to the pneumatic diagram in Table 8-8-6, Table 8-8-7, Table 8-8-8.
8-11
PRINTED DOCUMENTS ARE UNCONTROLLED
Vedere la pagina 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 224 225

Commenti su questo manuale

Nessun commento